Gas Analyzer Gas Detector Laser Dust Particle Counter Particle Monitor Laser Dust Particle Counter
Product introduction
1,Product name: Particle counter | 2. Product model: LR-CLJ-E3016 |
3, light source: semiconductor laser | 4, sampling flow: 2.83L/min |
5, particle size channel: six channel simultaneous counting | 6.power supply: AC220 50HZ |
7, size: 200*280*110mm | 8, weight: 2.6kg convenient to carry |
Gas Analyzer Gas Detector Laser Dust Particle Counter Particle Monitor Laser Dust Particle Counter
The product features
The technical indexes of the series of instruments all meet the requirements of jJF1190-2008 verification regulation issued by the State Administration of Metrology. The whole machine adopts the American microcomputer control processing technology, and can directly print the test results.It has many functions, high measurement accuracy, high speed, easy to carry and easy to operate.The instrument can measure the number of dust particles of various particle sizes at the same time, and can choose to observe the number and change of a particle size, which is very convenient for research, detection and evaluation of various clean environment.
Gas Analyzer Gas Detector Laser Dust Particle Counter Particle Monitor Laser Dust Particle Counter
Range of application
The series of products have been widely used in clean room testing;Filter site detection, leakage detection;Can monitor ultraclean workbench, HVAC system, computer room, the production of medical equipment, hospital clean operating room, car spraying, pharmaceutical and biological products, food hygiene, fine chemical industry, precision machinery and aerospace and other production and scientific research departments, is to implement the requirements of GMP pharmaceutical enterprises and their supervision and regulation department and electronic manufacturing enterprises preferred instruments.The series of instrument performance design advanced, stable and reliable quality.
Gas Analyzer Gas Detector Laser Dust Particle Counter Particle Monitor Laser Dust Particle Counter
Main technical parameters and performance:
1. Light source: all-semiconductor laser light source
2.83L/min(0.1 CFM /min),
3. Detection range :100 ~ 1,000,000 / at the same time in line with the new GMP standard A/B/C/D four levels of static and dynamic standard measurement/four standards switch measurement
Allow the dust concentration using the tested air to be 100 thousand pieces /2.83L
4. Particle size channel:
Standard particle size: 0.3 0.5 1.0 3.0 5.0 10.0 (μm) six grades, if other particle size requirements can be customized.
5. Six sizes of dust particles are detected at the same time, with four display modes:
(1) Real-time display: Real-time display of a certain particle size of the current detected particle size data.
(2) Select display: select to display the accumulated value of a certain particle size in the previous period.
(3) Process display: Display the real-time data of each particle size in the current test successively.
(4) Sequential selection: the cumulative value of each particle size in the previous period is displayed successively.
6. Display or print can convert particles contained in 2.83 l/min (0.1ft3) to particles contained in 28.3 l/min (1Ft3) or 1m3.
7. Sampling period: 1-10 (min) Net time: < 15 min
8. Working environment: temperature :10 ~ 35ºC Relative temperature: 20 ~ 75%RH
9. Measuring temperature and humidity range and accuracy: optional (1) Temperature: 0 ~ 50 ° C ±1 ° C (2) Humidity: 0 ~ 100%RH±5%
10. Sampling points (A) 2-7 points set sampling times per point (L) 2-9 times set
11 UCL report: in line with iso14644-1, GB50073-2001 standards
12. Working time: 8 hours Power source: AC220V±10% 50±2Hz Maximum power consumption :20W
(1) Built-in long-acting lithium battery: long time continuous measurement work for more than 8 hours, external charger can be quickly filled (& LT;4 hours) The power supply of the charger is 110V ~ 220V,50/60Hz
13. Data storage: can store 1000 sets of data (including particle size, data, environmental data, year, month, day, time), data will not be lost after power failure.And can view and print data at any time.
14 alarm setting: the instrument with level alarm function, can alarm the clean room 100, 1000, 10000, 100000, 300000 level exceeds the standard
15. Note: Built-in printer, automatic judgment of purification grade, such as power sampling head, sampling rack (temperature and humidity probe is an optional part)
16 calibration: JJF-1190-2008 dust particle counter measurement calibration specification can be traced back to the United States National Institute of Standards technology (NIST), our company has passed the national measurement standard assessment, can be traced back to Shanghai Institute of Measurement Technology can also be calibrated or third-party national measurement institutions for calibration
17. Weight :2.5kg. Dimensions :200×110×280