The smoke point meter is used to measure the smoke point of kerosene and jet fuel for lamps, that is, the maximum height of the smokeless flame when the sample is burned in a standard lamp under specified conditions, expressed in millimeters.
HZYD-3820 Kerosene Smoke Point Tester is designed and manufactured according to GB/T382-83 (91) "Kerosene Smoke Point Determination Method".
Kerosene and jet fuel smoke point wicks comply with GB/T382-2017 <<Determination method for smoke point of kerosene and jet fuel>>, ASTM D 1322-08 Standard ASTM D1322 Standard Test Method for Smoke Point of Kerosene and Aviation Turbine Fuel
, for the determination of the smoke point of kerosene and jet fuel for lamps.
The smoke point, also known as the smokeless flame height, refers to the maximum height of the smokeless flame when the sample is burned in a standard lamp under specified conditions, expressed in millimeters (mm). The sample oil burns in a standard luminaire, and the change in flame height is reflected on the background of a millimeter scale. When measuring, raise the wick to a flame with smoke, and then lower it to the point where the smoke tail just disappears. The height of the flame at this point is the smoke point of the sample.
Test Standard | GB/T382,ASTM D1322 |
Tested sample | Petroleum products and chemical products |
Test results | Determination of the smoke point of kerosene and jet fuel, that is, the maximum height of a smokeless flame when the sample is burned in a standard luminaire under specified conditions, expressed in millimeters |
Model No. | HZYD-3820 |