Description
CXYD-100 Petroleum Products Dynamic Viscometer (Vacuum Pressure Reduction Capillary Method)
Use and range
This instrument is designed according to the request of People's Republic of China standard GB/1841 "Polyolefin Resin Solution Viscosity Test Method". It is applicable for determining the viscosity of declaim solution of polyethylene and polypropylene under the set temperature according to GB/1841. This instrument is also applicable for determining the bitumen specimen dynamic viscosity under set temperature according to People's Republic of China industrial standard T0620 "Bitumen Dynamic Viscosity Test (Vacuum Pressure Reduction Capillary Method)" (idt: ASTM D2171, JIS and AASHTO T202).
Main technical parameters- Working voltage: AC220V±10%, 50Hz
- Constant water bath heating power: 600W
- Constant water bath working temperature: 60ºC
- Constant water bath temperature control accuracy: ±0.1ºC
- Mercury thermometer: bar type, division value 0.1ºC Temperature measuring range: 50-100ºC
- Constant water bath capacity: no higher than 20L
- Specimen quantity: capillary quantity one time: 3pcs
- Mixing motor: power 6W; rotation speed 1400r/min
- Ambient temperature: -10ºC~+35ºC
- Relative humidity: <85%
- Temperature sensor: industry platinum resistor, the division number is PT100
- Vacuum pressure reduction system: pressure digital display, automatic constant pressure, DC micro-vacuum pump (40W, 5L/min)
- Power consumption: no higher than 700W
- Vacuum pressure reduction capillary viscometer: one group including: 25, 50, 100, 200, 400, 400R, 800R