Usage:
Orthopaedics: osteopathy, diaplasis, nailing
Surgery: removing foreign body, cardiac catheter, implanting pace maker, interventional therapy, partial radiography, local photography, and other work.
Features
1. New integrated design, compact appearance, convenient operation
2. High-quality combined HF-HV X-ray generator, greatly reducing the X-ray irradiation.
3. Electric rotatable adjustable collimator
4. Unique hand-held remote controller, more convenient operation
5. Graphical colorfully LCD touch screen console interface, convenient operation, fashionable design
6. Toshiba image intensifier is employed, stable and reliable in quality, high-definition image and display system
7. With the perspective KV, MA automatic tracking feature, the image brightness and definition are automatically set at optimal function.
Configuration
No. | Item | Quantity |
1 | Main frame of C-arm | 1 |
2 | X-ray generator and high frequency inverted power supply | 1 |
3 | Toshiba 9'' image intensifier | 1 |
4 | Medical 1 Mega-pixels digital CCD camera | 1 |
5 | CCU | 1 |
6 | 23'' high-definition Medical LCD monitor | 1 |
7 | Imported minigroove grid | 1 |
8 | Electric adjustable collimator | 1 |
9 | Hand controller | 1 |
10 | Red light cross localization | 1 |
System Performance and Parameter:
Category | Project | Content |
Electrical Performance | High Frequency Inverter Power Supply | Output Power: 3.5kW Inverter Frequency:40kHz |
Automatic Fluoroscopy | Tube Voltage:40kV~110kV automatic adjustment Tube Current:0.3mA~4mA automatic adjustment | |
Manual Fluoroscopy | Tube Voltage:40kV~110kV continuous adjustment Tube Current:0.3mA~4mA continuous adjustment | |
Pulse Fluoroscopy | Tube Voltage:40kV~110kV continuous adjustment Tube Current:4.1mA~8mA continuous adjustment | |
Photography Tube Voltage, mA, mAs | 40 kV~110 kV 20-63mA 1-125mAs | |
film holder Size | 200mm×250mm(8″×10″) or 250mm×300mm(10″×12″) | |
X-ray Tube | X-ray Tube Special for High Frequency | Fixed anode Dual-focus, small focus: 0.3, large focus:1.5mm Thermal capacity: 35KJ (47KHU) |
Video System | Image Intensifier | TOSHIBA 9″ Image Intensifier |
CCD Video camera | Medical 1 Mega-pixels digital CCD camera | |
Monitor | Horizontal center resolution 1000/L, frequency for picture refresh is 75HZ | |
CCU main control | High Definition progressive output mode, 8 level Noise Reduction,8 image storage, Vertical Image, Horizontal image flip, positive or negative image | |
Structure | Directive wheel | Direction wheel 360°rotation, main wheel ±90°rotation |
C-arm | Electric Ascending & Descending of pillar movement: 400mm Forward and Backward Movement: 200mm Slip on orbit: 120°(+90°~ -30°)
|