Hot Sell Quick Response Molten Steel Hydrogen Probe

Min.Order: 100
Product origin: Cangzhou, Hebei, China
Infringement complaint: complaintComplaint
US$ 5 ~ 8

Description
Product Description
      Molten Steel Hydrogen Probes
HYDRIS could measure the hydrogen content of the molten steel in ladle, tundish and ingot mold directly. As the carrier gas,nitrogen passes through the probe, enters the molten steel liquid and siphons off the hy drogen from the molten steel. The process will be ended until the hydrogencontent is balance between the carrier gas and the molten steel liquid.
 
 
DesignationTemperature 
Range
Measuring RangeMeasurement accuracyResponse timeMeasured Rate
Hydrogen Probes1200-17500.4-13PPM0.1PPM28s-70s
98%
 
 
Working method

1. As the carrier gas, nitrogen is pushed by the pump in the circuit. It enters the molten steel liquid and forms bubbles after passing through the gas pipeline, measuring lance and the blowing pipe of the probe. The nitrogen bubble comes back to the gas cabinet by passing through the air plug of bell jar in the probe after absorbing the hydrogen from the molten steel liquid. The carrier gas will circulate in the circuit many times until the hydrogen content is balance between the carrier gas and the molten steel liquid.

2. 
The thermal conductivity tester of the pneumatic cabinet is able to analyze the hydrogen content in the carrier gas.
According to SIEVER law, the hydrogen content in the molten steel liquid could be calculated by the Hydrogen pressure.

Instructions

1. Check the inlet pressure of pneumatic cabinet for nitrogen and the service condition of loop filter. Make sure the blinking cursor in the lower right comer of instrument screen indicatesthe ready status.

2. Press the start button of the pneumatic cabinet or press the F2 button of the instrument, and the system begins to work. The hydrogen measuring lance blows out and the instrument screen shows the bolwing press constantly rises. About 10 seconds later, the yellow light of the pneumatic cabinet is on while the red light is still on.

3. Install the hydrogen probe carefully and screw up tightly, but prohibit to impact hardly. After 4-5 seconds, the green light is on while the red light is off. In the meanwhile, the yellow lightkeeps on. The blowing press on the instrument screen con stantly rises to 4500Pa. Yellow light is off while green light on, the system gets ready for the measurement.

4.It must be stopped blowing argon before measurement until the molten steel liquid has been quiet for about 1-2 minutes. It should be avoided to measure during the period of adding alloy into the liquid or measure the hydrogen content immediately after adding alloy.

5.Lift the hydrogen measuring lance carefully and insert it into the molten steel liquid while keepthe probe vertical to the liquid surface. The depth of the insertion is about 300-400mm. Hold the lance steady and keep balance to avoid of shaking. Prohibit lift the lance before finishing the measurement until the red light is on what represents the end of measurement(in the meanwhile the buzzer rings out in the pneumatic cabinet.

6. 
After measurement, take the probe out timely and carefully and check the contact block is clean or not. Put the hydrogen measuring lance away and do not leave the used probe connected with the lance in order to avoid of damaging the hydrogen measuring lance.
Company Profile

      Cangzhou Qi Run Detecting Instrument Co. Ltd. was founded in 1996, is a professional production company for metallurgical industry detection, sensor and instrument. The company covers an area of 50000 square meters, construction area of 25000 square meters,and has more than 200 employees. Qi Run owns many workshops such as matel product, refractory product, paper tube, equipment manufacturing, sampler sensors, research and other major production workshops.More than 95% of the parts have achieved self-control, which effectively reduces the cost and enhances the competitiveness of us.Now our company Qi Run has achieved the annual production capacity which are listed below: Converter sublance 500 thousand pieces per year, sampler 3 million pieces per year,sampler parts 5 million sets per year, thermocouple 60 million pieces per year, oxygen probe 100 thousand pieces per year. 
     
Our products are supplied to not only domestic steel corporations but also many other countries and regions such as USA,Japan,Australia,Thailand,Mexico,Columbia,Pakistan and so on.

Certifications

FAQ

Q: Are you Factory?
A: Definitely, we are a factory. 

Q: What Parameters should I provide to you for a proper quotation?
A: Application, pore size, dimension,, quantity, packing requirements, trade terms.


Q: How can you control your quality?
A: For each production processing, we have a complete QC system for the chemical composition and physical properties. After production, all the goods will be tested, and we have the quality certificates with ISO9001.


Q: What is your lead time?
It usually needs about 15- 20 days after receiving the PO.

Q: Do you provide Samples?
A: Yes, our factory provides samples for testing, pls send detailed parameters to us. (if customized, withdrawing is best) The samples will be charged or free according to the specific requirements.

Q: What are the Payment Terms?
A: We accept T/T, L/C, etc.


Q: How can I get samples?
A: We are honored to offer you the samples.

Q: What about your Packing?
A: We provide carton boxes, pallets, etc. OEM packing is available too.

Q: Do you have R&D services?
A: Yes, we have an R&D department to assist you in new product development.


Q: Does your company accept customization?
A: We accept OEM service.

Q: Can we visit your company?
A: Yes, for sure, you are welcome to visit our company.

Scroll to Top