AA2610 Atomic Absorption Spectrometer Device for Metal Analysis Graphite Furnace Spectrometer

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 1000 ~ 3000

Description

AA2610 Atomic Absorption Spectrometer Device for Metal Analysis Graphite Furnace Spectrometer
Performance index
1.Optical system
Wavelength range 190nm -- 900nm
Monochromator anti-aberration C-T monochromator
Spectrum bandwidth 0.2, 0.4, 1.0, 2.0nm, four automatic selectable
Wavelength accuracy ±0.30nm
Wavelength repeatability 0.20nm
Baseline drift static ±0.002A/30min Dynamic ±0.003A/30min
Detector photomultiplier tube
When the resolution is 0.2nm spectral bandwidth, the half-width of the measured spectral line does not exceed 0.2nm±0.02nm

2. Atomization system
Sensitivity (Cu) Mass concentration 2.0ug/ml absorbance greater than 0.300Abs
Detection limit (Cu) 0.006μg/ml
Burner 100mm metal titanium burner, air cooled premixed type
Precision RSD≤0.8%
Sprayer Efficient glass atomizer
Atomization chamber Corrosion-resistant material atomization chamber
Position adjustment Flame burner suitable height and front and rear position can be adjusted, one minute to complete flame/hydride replacement
Safety measures with a variety of automatic safety protection functions, acetylene leakage warning, closing system

3. Analytical method
Measuring method Air-acetylene flame method, hydride generator atomic absorption method, oxygen-enriched air-acetylene flame method
Concentration calculation method Standard curve method (1-3 curve), automatic fitting, standard addition method
Repeated measurement times 1~20 times, calculate the average, give the standard deviation and relative standard deviation
Result printing parameter printing, data result printing, graphic printing, summary report
Software environment Windows 98/2000/XP

Instrument parameter
1. Host machine
Power supply AC 220V/50HZ
Power 150VA
Weight 70kg
Volume 1000mm(L)×350mm(W)×390mm(H)
Working environment temperature 10~35ºC
Working humidity ≤80%

Standard configuration
1. Host
2. Low noise oil free air compressor
3. Operate the software
4. Come with accessories
5. Installation and commissioning
6. Customer on-site training methods
7. Packing and transportation in wooden cases
Special note: To use this model, you should purchase a local PC, if you need to print experimental data, you should also purchase a printer. If it is not convenient to buy a personal computer or printer, we can also buy it on behalf of our company.

Four. Optional accessories
1. Hydride generator
2. Quartz concentrator
3. Oxygen enrichment device

 
Product Tag:
Related categories:
Scroll to Top