5 different types for below each series available: Neonate clip, Pedia clip, Adult soft, Pedia soft, Neonate wrap.
Reusable SpO2 Sensor Probes
The product is indicated for reuse to collect and transmit signal of SpO2 and pulse rate of patient, with compatible monitors or oximeters. The product is only used by experienced medical professional or under their directions.
Operation principle, function and structure
The patient sensor end of the product is clamped the patient's index fifinger with (Thumb, middle fifinger or others, great toes), and the device connector is connected to compatible monitors or oximeters with SpO2 extension cord. The product adopts 660nm red light and 905nm Infrared light as incident light source, tests light intensity through tissue to obtain SpO2 value as well as pulse rate.
Specification
Measurement wavelengths and Output Power: Red:660±5 nm @ 2.0mw maximum average
Infrared: 905 nm±5 nm @ 2.0 mw maximum average Oxygen Saturation Declared
Accuracy Range: ±3% of 70%~79% spo2; ±2% of 80%~89% spo2; ±1% of 90%~100% spo2;
Pulse Rate Range: 30~245 BPM;
Accuracy: ±3%bpm of 150 bpm~245 bpm; ±2 bpm of 60 bpm~149 bpm; ±1% bpm of 30 bpm~59 bpm