Biobase Automated Blood Testing Cbc Cell Counter Medical 3-Part Hematology Analyzer

Min.Order: 1
Product origin: Jinan, Shandong, China
Infringement complaint: complaintComplaint
US$ 1800

Description
 

Biobase Medical Blood Analyzer 3 Part Hematology Analyzer Hemocytometer

Features

 

Features:
 

.Throughput: 60T/H

.8 inch touch screen

.20 parameters + 3 histograms

.3 counting modes

.100,000 sample results

.Support LIS and external printer

 
Recommended hot products

 


Technical Parameters

Model

BK-3100

Throughput

60 samples/hour

Principle

Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB

Channels

2

Parameters

21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDW,PCT, P-LCR, P-LCC)

Histograms

3 histograms (including WBC, RBC, PLT histograms)

Sample Volume

Whole blood: 10μl, Prediluted: 20μl

Calibration

Manual, auto calibration

Counting Modes

Venous, capilary, prediluted

Printing Model

Auto print, manual print

LIS

Support Bi-directional LIS

Language

Chinese, English, Spanish, French, Russian, Ukrainian

Reagent

Diluent 20L, Hemolytic agent 500ml, Probe cleanser: 50ml

Data Input

9.7-inch touch screen, mouse, keyboard (optional)

Output

Internal printer, support external printer

Printer paper

57*30mm

Interface

4 USB ports, network port,RS232 serial port.

Storage

100,000 results with histogram

Unclog

High voltage burn, reverse high-pressure flush

Work Temperature

15-30°C

Power Supply

100~240 AC, 50/60Hz, 150VA

Relative Humidity

30~85

Atmospheric Pressure

70.0kPa-110.0kPa

Carry-over rate

Parameter

WBC

RBC

HGB

PLT

CV

≤ 1.5%

≤ 1%

≤ 1%

≤ 3%

Background

Parameter

WBC

RBC

HGB

HCT

PLT

Background

≤ 0.3×109/L

≤ 0.03×1012/L

≤ 1 g/L

≤ 0.5%

≤ 10×109/L

Linearity

Parameter

Measurement range

CV

WBC

(0.5~10.00×109)/L
(10.10~99.9×109)/L

(±0.5×109)/L
±5%

RBC

(0.05~1.00×1012)/L (1.01~9.99×1012)/L

(±0.05×1012)/L
±5%

HGB

(2~70g)/L
(71~300.0g)/L

(±2.0g)/L
±2%

PLT

(10~100×109)/L
(101~999×109)/L

(±10.0×109)/L
±8%

Precision

Parameter

Whole blood (CV)

Pre-diluted(CV) 

Measurement range

WBC

≤ 2.0%

≤ 4.0%

(4.0-15.0×109)/L

RBC

≤ 1.6%

≤ 3.0%

(3.5-6.0×1012)/L

HGB

≤ 1.6%

≤ 3.0%

(110.0-180.0g)/L

MCV

≤1.0%

≤2.0%

(70.0-120.0)fL

PLT

≤ 4.0%

≤8.0%

(150.0-500.0×109)/L

External Size(W*D*H)

295*467*418mm

Packed Size(W*D*H)

Instrument: 404*564*616mm; Reagent: 480*480*480mm

Net weight

19 KG

Gross weight

Instrument:31KG; Reagent::25Kg

Packaging & Shipping

Company Profile

 

Factory Views:
Founded in 1999, BIOBASE Group having over 2000 employees (till August, 2018), BIOBASE Group is a new high-tech enterprise focusing on development, production and management of scientific equipment, In Vitro Diagnosis (IVD) instrument and reagent.

 
BIOBASE Group is specialized in products of 8 areas including medical diagnosis, biosafety protection, disinfection and sterilization, water purification system, infant care products, cold chain products, software products, clean room project.

 
Contact Us

 




 
Product Tag:
Related categories:
Scroll to Top