SY-B143i Semi-automatic Biochemistry Chemistry Analyzer Blood Chemistry Analyzer for Hospital and Clinic
Features:
1. Analytic Modes: kinetic, fixed time, end-point, absorbance 2. Wavelengths: 7 standard filters: 340, 405, 450, 505, 546, 578, 630 nm, 2 more filter optional 3. Full open reagent system 4. Items can be programmed, number of items not limited 5. Display: LCD with high brightness 6. Input: Touch screen and keyboard & mouse operation 7. Real time graph display and print 8. Lamp saving feature, longer lamp lifespan
Specification:
Assay Method Reaction rate, end point, two-point, double wavelengths, coefficient, multiple standards, blank reagent, blank sample, absorbance at end point, QC, QC graph and turbidity Temperature System PELTIER temperature control component, 25ºC, 30ºC, 37ºC, accuracy: ±1ºC Light Source 6V 10W halogen lamp of more than 2000 hours lifespan GML Precision ±2 nm FWHM ≤10 nm Stray Light ≤0.5(absorbance≥2.5) Absorbance Range -0.214-3.000 Absorbance Linearity ±1% Absorbance Repeatability ±0.001 Precision ≤1% Carry Over ≤1% Resolution 1mAbs/h Stability ≤3mAbs/h Flow Cell Permanent quartz flow cell, minimal volume is 32uL, and disposable flow cell is selectable Sampling System Pressure auto releasing pump ensure the accuracy of suction volume