ISO 105-G03 Color Fastness Tester to Ozone

Min.Order: 1
Product origin: Wuxi, Jiangsu, China
Infringement complaint: complaintComplaint
US$ 5000 ~ 20000

Description

 ISO 105-G03 Color Fastness Tester To Ozone 



 
Model : FZ-109
Standards : GB/T7762-2003 GB/T2951.21-2008 GB/T 11206-2009  AATCC 109 ISO 105-G03

This test equipment is used to simulate the effect of atmosphere ozone environment to textiles and fabrics, it also can quickly identify and evaluate the anti-ozone aging performance and the protective effectiveness of anti-ozone agents.

Specifications 
Test space150L
Inner dimensionsW500 X H600 x D500mm
Outer dimensionsW1100 X H1700 x D1050mm
Approx. net weightApprox. 220kg
Power supplyAC220±5%V /50±0.5Hz
Temperature range 0 ºC -80 ºC
Temperature fluctuation ± 0.5 ºC ± 2.0 ºC
Ozone concentration
accuracy
10~1000pphm ± 10%
Sample holder lifting frequency 2R/min (optional)
Air flow velocity12-16mm/s
Test gas flow rate20-70L/min
Sample holder speed 360 degree rotation of the sample holder (speed 1-3 revolutions per minute or optional adjustable)
Humidity range20-98%

Constructions
Construction typeIntegration
Material made
Linear SUS304
Outer SUS304 or paint baking finish
Heat Insulation Rigid polyurethane foam+aluminum silicate insulation cotton double protection
Chamber gate
Opening modeSingle door, left open
Door /window frameTwo silicone rubber sealing strips and anti-condensation electric heating device.
Configurations Equipped with observation window W300xH450 mm and lighting
Sample holder requirements The fixture should be made of materials that are not easy to decompose ozone, such as aluminum. Optional dynamic test sample holder.
Control panel configurations1. Controller display screen
2. Power switch
3. Illumination button,
Other configurations1. Equipped with a stainless steel SUS # 304 storage rack turntable two φ 1 50mm lead hole with hole cover and silicone rubber plug.
3. Move 4 casters
4. Four horizontally fixed foot stands
5. Rear of the box 


Test method
The test methods can be divided into "static" and "dynamic" according to the test methods and standards. Static refers to placing the sample in the test chamber for testing after being stretched, while dynamic refers to placing the sample on the fixture of the test chamber and conducting the test while stretching. The tensile deformation amplitude is about 5-45% of the length of the tested sample itself.
Place the sample in a certain concentration of ozone environment until a certain color change occurs or after a specified time, compare it with the standard chart