Medical Serum Electrolytes Open System Machine ISE Electrolyte Analyzer

Min.Order: 1
Product origin: Nanjing, Jiangsu, China
Infringement complaint: complaintComplaint
US$ 800 ~ 2000

Description

Medical Serum Electrolytes Open System Machine ISE Electrolyte Analyzer

Features of Electrolyte Analyzer
Parameter: K, Na, Cl, iCa, nCa, TCa, PH, Li, TCO2, AG
Automatic electric potential tracking and correcting software to ensure the stable performance
Automatically detect and filter tiny bubbles to avoid clog and ensure accurate measurement
Real-time diagnostic of system working status, waste liquid automatic detection and alarming
Automatic calibration and two-point correction to adjust slope and intercept
Wave theory flushing method and direct flushing pipe method to avoid block and crossed contamination
Power failure protection to avoid data losing, data storage could be extended to more than 20000
Available to power off any time, therefore reduce reagent consumption, suitable for any hospitals

Parameter of Electrolyte Analyzer: 
Measuring time≤25~90s(Type A~Type I,Time for sampling, measuring, 
washing and printing)
ParametersK, Na, Cl, iCa, nCa, TCa, PH, Li, TCO2, AG
Sample size60~300μl(Type A~Type I,optional)
Applicable specimensblood serum,blood plasma,whole blood,cerebrospinal fluid and dilute urine
Data storageup to 10,000 test results
Automatic samplingNo
DisplayLCD
Operating modeYES\NO button
Printinternal graphic thermal printer
Alarm for waste liquid overflowYes(optional)
Alarm for standard liquid levelYes(optional)
Air bubble test Yes(optional)
Barcode scanningNo
CPU16bit
communication interfaceRS232
Online softwareNO
Working conditionTemperature:  5ºC~35ºC Relative humidity ≤80%
PowerAC220V±20;50Hz±1Hz
Weight12kg(Host machine)