Medical hanging tower is indispensable multi-functional medical equipment for modern operating room. It is mainly composed of basic frame, connecting tube, rotating joint, rotating arm, box body, instrument platform and infusion stand.
The cabinet is equipped with a lifting and lowering liquid rack, a multi-layer instrument platform, a drawer, a medical gas terminal, a universal power socket, a protective grounding terminal, a network socket, a telephone socket, and can carry a variety of medical instruments and equipment.
The instrument platform features a new rounded bumper design that allows each platform to adjust position independently. The cabinet can be rotated by 340° and equipped with anti-drift damping device for precise rotation and positioning. The rotating arm can be arbitrarily positioned within a certain radius by pneumatic brakes, enabling medical personnel to operate various medical equipment handily.
The main body of the hanging tower is made of high-strength aluminum alloy profile which is formed once. The appearance treatment adopts imported anti-UV and anti-ozone raw materials. It is cured by high-pressure electrostatic spraying and high-temperature baking to form the surface of the composite material. The color remains unchanged for a long time. The agent is cleaned to completely eliminate the pollution.
The main technical parameters
1. Air intake pressure: | 0.4Mpa-1Mpa |
2. Negative pressure suction: | 0.05MPa±0.01Mpa |
3. Gas source terminal performance: | secondary sealing, more than 20,000 times of plugging and unplugging |
4. Hermeticity: | 1Mpa, no more than 0.296ml/min |
5. Input air pipe interface: | 6YC ball cap connector |
6. Input power: | 0~220V, 50Hz |
7. Power socket: | 8 international universal power sockets |
8. Maximum output power: | 4400W |
9. Maximum rotation angle of the box: | 330° |
10. Maximum rotation angle of the rotating arm: | 340° |
11. Platform specifications: | 350 × 420mm |
12. Platform carrying weight each platform can carry 25KG | |
13. Total load weight 100KG (up to 4 platforms can be configured) |
Precautions
1. Full implementation of products (GB 9706.1-2007) "Medical Electrical Equipment Part 1: General Requirements for Safety"
The type of electric shock protection belongs to Class I equipment. The protection against electric shock not only relies on basic insulation, but also adopts additional safety measures connected with the protective grounding wire, so that the accessible metal parts will not be charged even if the basic insulation fails.
2. The protective grounding wire supplied to this product must be safe and reliable, and should be taken separately from the grounding bus bar.
3. The medical valve provided to the product must be equipped with a medical maintenance valve at the end to facilitate installation and maintenance.
4. The rotation angle of the rotating arm and the box of this product is limited. Do not force the limit device to avoid damage to the equipment.
5. The instrument platform load capacity of this product is limited. Please refer to the technical parameters of the load capacity to place the medical instrument.
6. The surface of this product is sprayed with anti-UV and anti-ozone materials. Do not hit or knock with sharp objects to avoid damage to the coating.
7. In order to ensure the load-bearing factor safety load requirements, the base frame and the roof of the equipment must be reliably connected by expansion bolts to ensure that they can withstand more than 300KG.
8. The network power supply should have the quality used in a typical commercial or hospital environment.
1. Conditions that should be met before the product is installed
1.1 The location of the device installation has been determined. And will not interfere with other devices.
1.2 Each gas pipeline has been laid to the equipment installation and equipped with a medical maintenance valve.
(The model of medical maintenance valve is 6YC, the interface thread is M14×1.5)
1.3 The power cord has been reserved (single-phase L+N+PE) with a specification of 3×2.5mm2
1.4 The cable and telephone line should be prefabricated with a crystal head.
1.5 The device does not have a power switch. Users need to install an air switch on the wall of the room.
The air switch has a specification of 20A and must be able to cut both the phase and neutral lines of the power supply at the same time.
Medical maintenance valve legend
2. Install the base frame
2.1 With the base frame installation position as the center, open a round hole in the ceiling with a diameter of Φ360mm.
2.2 Fix the base frame to the concrete floor with M12×100 expansion bolts.
(Supporting drill bit 16mm, matching wrench 20)
2.3 The height of the base frame is 150mm from the bottom of the ceiling.
3. Install the connecting pipe assembly
3.1 Install the top cover, flange, and round nut onto the connecting tube assembly in the order shown.