Technical Characteristics and Parameters | |
Syringe specification | 5, 10, 20, 30, 50(60)ml Disposable sterilized syringe |
Preset value | (0-9999.9)ml |
KVO speed | (0.1-5)ml/h User settings; The increment is 0.1ml/h |
Injection speed | 5ml syringe: (0.1-100)ml/h 10ml syringe: (0.1-300)ml/h 20ml syringe: (0.1-600)ml/h 30ml syringe: (0.1-800)ml/h 50ml (60ml)syringe: (0.1-1800)ml/h |
Bolus speed | 5ml syringe: (50-100)ml/h 10ml syringe: (50-300)ml/h 20ml syringe: (50-600)ml/h 30ml syringe: (50-800)ml/h 50ml (60ml) syringe: (50-1800)ml/h |
Purge speed | 5syringe: 100 ml/h 10 syringe: 300 ml/h 20 syringe: 600 ml/h 30 syringe: 800 ml/h 50 (60ml) syringe: 1800 ml/h |
Bolus | 1-10ml |
Capacity precision | 0.1-5ml/h ±5%; >5ml/h ±2% |
Speed unit: | ml/h, mg/min, mg/h, mg/24h, mg/kg/min, mg/kg/h, mg/kg/24h, ug/min, ug/h, ug/24h, ug/kg/min, ug/kg/h, ug/kg/24h, u/min, u/h, u/24h, U/kg/min, u/kg/h, u/kg/24h, kU/min, kU/h, kU/24h, kU/kg/min, kU/kg/h, kU/kg/24h |
Occlusion pressure | (40.0-100.0)±20kpa (High, medium and low gears can be adjusted) |
Alarm type | AC power down, syringe falling off, installation error, speed anomaly, low power, power depletion, drugs to be empty, pushing to empty, Occlusion, and completion |
Alarm volume | Low, medium and high levels can be adjusted Low end is no less than 45dB and high level is no less than 60dB |
Key tone | Selecting open or close |
Drug library | It is internally provided with 50 commonly used drugs |