Ventilation modes: | IPPV; A/C; SIMV; |
Ventilator parameter ranges | |
Flow meter | O2(0.1-10L/min) |
N2O(0.1-10L/min) | |
AIR(0.1-10L/min) | |
Rapid oxygen supply | 35L/min-75L/min |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~ 30 cmH2O |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 60 cmH2O |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Oxygen concentration | 15 % ~ 100 % |
Graphical display: | Oxygen concentration |
Size Wooden case packing size:L 560*W 820*H 1450mm , GW:104KG ; NW:64KG Alarm and protection | |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit | Show on screen Show on screen |
Working conditions | |
Gas source | O2,N2O |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 40VA |
Vaporizer |
Anesthesia gas | Adjustable scope % (volume percentage) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
Sevoflurane | 0 ~ 8 |