Model | BI-3000 | BI-1000 | BI-800 |
Canopy Specifications | |||
Mattress (canopy) | 450mm | 450mm | 450mm |
Access Port | 6 (2 Iris ports) | 6 (2 Iris ports) | 4 (2 Iris ports) |
Self-sealing Tube Ports | 4 | 4 | 2 |
Front Access Panel | 826x310mm | 826x310mm | 826x310mm |
Mattress Tray Size | 630x360mm | 630x360mm | 630x360mm |
Technical Specifications | |||
Temperature Mode (Servo Controlled) | air, infant | air, infant | air |
Air Temperature Adjustment Range | 20°C~37°C, 37~39°C | 20°C~37°C, 37~39°C | 20°C~37°C |
Infant Temperature Adjustment Range | 30°C~37°C, 37~38°C | 30°C~37°C, 37~38°C | N/A |
Infant Sensor Accuracy | ≤0.3°C | ≤0.3°C | N/A |
Relative Humidity Display Range | 5%~99% | N/A | N/A |
Temperature Fluctuation | ≤0.5°C | ≤0.5°C | ≤0.5°C |
Uniformity of Mattress Temperature | ≤0.8°C | ≤0.8°C | ≤0.8°C |
Time to Warm Up | ≤30 min | ≤30 min | ≤30 min |
Noise Level within Canopy | ≤55dB | ≤55dB | ≤55dB |
Alarms | |||
Power Failure Alarm | Yes | Yes | Yes |
Fan Failure Alarm | Yes | Yes | Yes |
Sensor Failure Alarm | Yes | Yes | Yes |
Air Temperature Deviation Alarm | ±3.0°C | ±3.0°C | ±3.0°C |
Infant Temperature Deviation Alarm | ±1.0°C | ±1.0°C | N/A |
Air Temperature Exceeds set Limit Alarm | ≤38.0°C; ≤40.0°C | ≤38.0°C; 40.0°C | ≤38.0°C |
Infant Temperature Exceeds set Limit alarm | ≤38.0°C; ≤39.0°C | ≤38.0°C; 39.0°C | N/A |
System Failure Alarm | Yes | Yes | N/A |
Mute Alarm Function | Yes | Yes | N/A |
Operational Environment | |||
Temperature within Operational Environment | +20°C~+30°C | +20°C~+30°C | +20°C~+30°C |
Relative Humidity within Operational Environment | 5%~99% | 5%~99% | 5%~99% |
Velocity of Air Flow within Operational Environment | <0.3m/s | <0.3m/s | <0.3m/s |
Other Specifications | |||
RS-232 connector | Yes | Yes | N/A |
Tilt of Infant Bed (Adjustment Range) | ±10° step-less | ±10° step-less | N/A |
Maximum Load of Mattress Tray | 5kg | 5kg | 5kg |
Maximum Load of IV Stand | 2kg | 2kg | 2kg |
Maximum Load of Cabinet | 6kg | 6kg | N/A |