Description
1. Product brief introduction:1) SF6 Multi function tester can measure Decomposition, Dew point, purity in one time, it can greatly save gas, it's the best instrument to protect environment and reduce pollution. Meanwhile, it can reduce user's workload to improve work efficiency.
Why the the tester has the excellent proformance, the reason is it uses all the best sensor in the world: Dew Point sensor from Vaisala Finland, Purity sensor from HNP German, Decomposition sensor from dedicated SF6 Dekang the United States.
3) The instrument has a large LCD display to show various real-time parameters, the whole fool operation, massive information storage, rechargeable battery for AC / DC purpose. It usually got a good praise from users because of high-tech and environmental protection and good proformance.
Function1) Graphical menu, color display
2) Electronic flow display and electric quantity display
3) Measurement data can be stored, printed
4) Support for data communications.
5) Support file system.
6) Self diagnosis function
7) Easy operation, convenient maintenance
8) Data management system software
9) AC / DC purpose, built-in lithium battery
10) Small volume, light weight for portable
Parameters1)Display: Colour large LCD
2) Power Input: 220VAC± 10%50/60Hz, AC/DC purpose, overcharge
Protection, built-in 3500mAh lithium battery
3) Working pressure: 0-1.0MPa
4) Gas flow velocity: No effection, Electronic mass flowmeter
5) Output Interface: RS232
6) Operation Environment: -20-+60
7) Humidity: 0-100%RH
8) Size: 250*150*300mm3
9) Weight: Approx 5kgs
SF6 Dew Point:
1) Test Range: Dew Point -80~+20C (Support PPMV...)
2) Accuracy: ± 0.4C(The sensor outputs frost point when Dew Point
Temperature is below 0C)
3) Response time: 63% 90%
+20- -20CTd 5s 45s
-20- -60CTd 10s 240s
4) Resolution: Dew point 0.1C, or 0.1PP
5) Repeatability; : ± 0.4C
6) Gas flow: SF6 0.5-0.9L/min adjustable
7) Probe protection: Stainless steel filter
SF6 Purity:
1) Test Range: 0-100%SF6;
2) Accuracy & Repeat: ± 0.4%, unrelated to flow;
3) Response time: 90%, 15s
SF6 Decomposition
1) H2S & SO2:
Measuring range: 1-200ppm(Standard)
Mini testing: ≤ 0.1ppm
Acurracy: ± 0.5%
Stability: 0~200ppm
Repeatability: ≤ 2%
2) HF: 0-10ppm
Purity | Test range | 0%~100% |
Accuracy | ±0.5% |
Test time | <2 min. |
Dew point | Test range | -80 C~+20 C |
Accuracy | ±0.5C (-80C~-60C) |
Response time (+20C) | 63% is 5sec,90% is 45sec (-60C~+20C) 63% is 10sec,90% is 24sec (+20C~-60C) |
H2S | Test range | 0~200ppm |
Testing min Value | ≤0.1ppm |
Accuracy | ±0.5% |
Stability | 0~200ppm |
Repeatability | ≤2% |
SO2 | Test range | 0~200ppm |
Testing min value | ≤0.1ppm |
Accuracy | ±0.5% |
Stability | 0~200ppm |
Repeatability | ≤2% |
HF | Test range | 0~20ppm |
Min check amount | ≤0.01ppm |
Accuracy | ±0.5% |
Stability | 0~20ppm |
Repeatability | ≤2% |
CO | Test range | 0~1000ppm |
Testing min value | ≤1ppm |
Accuracy | ±0.5% |
Stability | 0~1000ppm |
Repeatability | ≤2% |
Accessories Sr no. | Description | Quantity |
1 | SF6 Analyzer (Host) | 1 | unit |
2 | Portable aluminum alloy suitcase | 1 | unit |
3 | Test tube with valve for gas Input | 1 | unit |
4 | Soft tube for gas output | 1 | unit |
5 | Power supply cable | 1 | unit |
7 | Metal joint set box (for optional) | 1 | Set |
8 | Calibration report | 1 | unit |
9 | Operating manual | 1 | unit |