Description
GF3021 Laboratory electrical calibration Device for power transducer and power meter
GF3021 Laboratory electrical calibration Device is suitable for power grid company measuring and testing center, power supply company and power plants measuring department, national levels measuring and testing institutions, also suitable for railway, petroleum, chemical industry and large industrial and mining enterprises, etc.
Introduction
GF3021 Laboratory electrical calibration Device by program-controlled three-phase standard power resource, DC standard power resource, three-phase multi-function meter and the meter with standard and management software of the computer. The technical index device accord with country and industry related standards and verification regulation. Standard resource meter output, safe and reliable, easy to use, standard meter precise meter, the function is all ready, a meter is multi-purpose. The source of computer meters and verification system, to have corresponding output information is power inspection meter can realize the automatic verification, and meet the requirements of the rules are verification report.
FUNCTIONS AND FEATURES
All kinds of electric measurement transducer can be checked, including AC/DC voltage transducer, AC/DC current transducer, frequency transducer, phase transducer, single/ three-phase AC active power transducer, and 3-phase reactive power transducers.
Check all kinds of electric measurement indicating meter, including AC/DC voltmeter, AC/DC ammeter, frequency meter, phase meter, single three-phase acjige active power meter, three-phase ac reactive power meter, synchronous meter, etc.
Test single-phase, three-phase electronic, mechanical watt-hour meter or energy meter/ kWh meter error.
Calibrate AC sample device, RTU, measurement device error.
5The built-in electric measurement transducer, electric measurement instrument and meter instructions of verification procedures, can fully automatic or semi-automatic for verification, and save 1000 group check data.
It can be used as voltage source, current source and power source with high precision, and it is a high stability standard resource.
8-inch big screen color display and English interface.
For the software calibration, you don't need to open the case, it's stable and reliable.
Voltage output terminal with short circuit, current output terminal open protection and power amplifier overheating protection function.
With automatic failure detection function, shows fault part, the convenience users check line.
With USB port, it can connect computer for data management or controlled by PC.
PARAMETERS
Electrical parameters |
---|
Accuracy class | 0.05%, 0.1% |
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60Hz |
Communication port | USB, RS232, RS485, LAN |
AC Voltage output |
---|
Range(U1,U2,U3) | 50V, 100V, 200V, 400V, 600V |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Current output |
---|
Range(I1,I2,I3) | 0.5A, 1A, 2.5A, 5A, 10A, 20A |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Power output |
---|
Active output stability | 0.01%RG/1min |
Reactive output stability | 0.02%RG/1min |
Active accuracy | 0.05% RG |
Reactive accuracy | 0.1% RG |
Frequency output |
---|
Adjustment range | 45-65Hz |
Adjustment resolution | 1Hz, 0.1Hz, 0.01Hz and 0.001Hz |
Resolution | 0.001Hz |
Accuracy | 0.002Hz |
Power factor output |
---|
Adjustment range | -1 to 0 to +1 |
Adjustment resolution | 0.0001 |
Resolution | 0.0005 |
Phase output |
---|
Adjustment range | 0°-359.999° |
Adjustment resolution | 10°, 1°, 0.1°, 0.01° |
Resolution | 0.001° |
Accuracy | 0.05° |
Harmonic configuration |
---|
Times | 2 to 31 |
Content | 0-40% |
Phase | 0°-359.999° |
Configuration error | (10% RD + 0.1%), RD refers to the configuration value of harmonic contents |
DC Voltage output |
---|
Range | 75mV, 75 V, 150 V, 300 V, 500V, 1000 V |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1 min |
Distorting | ≤0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
DC Current output |
---|
Range | 0.5 A,1A, 2.5 A, 5 A, 10A, 20 A |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1min |
Distortion | ≤ 0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
Energy Error |
---|
Active error | 0.05% RG |
Reactive error | 0.1% RG |
DC Input Voltage Measurement |
---|
Range | 0 to ±20V |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
DC Input Current Measurement |
---|
Range | 0-20mA |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
Mechanical parameters |
---|
Dimensions (W×H×D) (mm) | 460x430x185 |
Weight (kg) | 20 |
Environmental conditions |
---|
Working temperature | 0°C to 40°C |
Relative humidity | ≤85% |