Single Phase Energy Meter Test Bench with Voltage Source and Current Source

Min.Order: 1
Product origin: Beijing, China
Infringement complaint: complaintComplaint
US$ 8858

Description
GF102 Single Phase Energy Meter Test Bench With voltage source and Current Source

GF102 portable single phase energy meter testing bench used to test single phase energy meter. Adopt mature signal synthetic and power amplifier technology. High precision voltage and current source output. Widely used in lab and field. And then as single phase standard source or single phase phantom load source.
GF102 portable single phase energy meter testing bench can automatically test energy meter error, and semi-automatic test energy meter error, test quickly, 7inch touch TFT color LCD display.
Features
Meter under testing: 1P2W
Testing mode: Active power, reactive power.
Test 3PCS or 2PCS meter synchronously.
Starting test and creeping test.
Reference standard and power source integrated, easy to carry.
Measuring mechanical meter and electric meter.
Operating by manual keyboard or PC software
With shortcut key of normal test point, speed up testing
Internal large capacity storage device for mass memory
Local checking the testing result
Overload, short circuit, open circuit protection
With self-checking function
Parameters
Electrical parameters 
Accuracy class0.05%, 0.1%
Power SupplyOne Phase AC 100-265V, frequency 50/60Hz.
AC Voltage Output 
Range57.7V, 100V, 220V, 380V; max 500V
Adjustment range(0-120)%RG(1)
Adjustment fineness0.01%RG, 0.1%RG, 1%RG, 10%RG as optional.
Stability0.01%/120s
Distortion0.1% (Non-capacitive load)
Output loadmax 25VA
Measuring accuracy0.05%RG
AC Current Output 
Range200mA, 1A, 5A, 20A, 100A; max 120A
Adjustment range(0-120)%RG
Adjustment fineness0.01%RG, 0.1%RG, 1%RG, 10%RG as optional.
Stability<0.01%/120s
Distortion≤0.1% (Non-capacitive load)
Output loadmax 50VA
Accuracy0.05%RG
Power Output  
Active power output stability<0.01%RG/120s
Reactive power output stability<0.02%RG/120s
Active power measuring accuracy 0.05%RG
Reactive power measuring accuracy0.1%RG
Phase Output  
Output adjustment range0°-359.999°
Output adjustment fineness10, 1, 0.1, 0.01 as optional.
Resolution0.01°
Accuracy0.05°
Power Factor 
Adjustment range-1 ~ 0 ~ 1
Resolution0.0001
Measurement accuracy0. 0005
Frequency Output  
Adjustment range40Hz-70Hz
Output adjustment fineness5Hz, 1Hz, 0.1Hz, 0.01Hz as optional.
Resolution0.001Hz
Accuracy0.005Hz
Voltage /Current/Harmonic Setting 
Harmonic number2-51times
Harmonic content0-40%
Harmonic phase0-359.99
Harmonic setting accuracy(10%±0.1%)RD(2)
Power Energy Measurement Error 
Active power energy0.05%RG
Reactive power energy0.1%RG
Power Pulse Output  
Power pulse typeactive pulse, reactive pulse
Active power pulse output 5V,  10mA
Power Pulse Input 
Energy pulse typesupport active and reactive pulse, the highest frequency power pulse input is 180K.
Mechanical parameters 
Dimensions (W×D×H) (mm)455x440x150
Weight (kg)12
Environmental conditions
Ambient temperature-10°C to +40°C
Relative humidity35%-85%
(1) RG means range, the same as below;
(2) RD means the setted harmonic content, harmonic can be a single output, also multiple output.
Product Tag:
Related categories:
Scroll to Top