Defibrillator: | |
Output | Biphasic Truncated Exponential |
Energy Sequence | 150, 150, 200J |
Charge Time | 8 sec. to 150J; 10 sec. to 200J (New Battery) |
Analysis Time | 9 sec. |
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery | |
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks | Less than 35 seconds |
The maximum time from initially switching power on to readiness for discharge | Less than 25 seconds |
Audible Prompts | 22 audible prompts |
Visual Prompts | LED prompts |
Controls | Two buttons-ON/Off, Shock |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω |
The Maximum Voltage | 1200±50V |
Output disabled when the patient impedance is outside limits | 20Ωto 200Ω |
Physical: | |
Dimensions | 303*216*89mm |
Weight | 2.0kg |
Operating Temperature | 0°C to 40°C |
| Relative humidity between 30% and 95% (non-condensing) |
Storage Temperature (without battery) | -20°C to 55°C |
Storage Humidity (without battery) | Up to 95% (non-condensing) |
Packaging size: | 43x25x32cm |
G.W. | 4KG |