TA2100 Wideband Power Calibrator

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 77500 ~ 79600

Description
TA2100 wide frequency power calibrator 

1. Summary

TA2100 Is a wide range, high precision, high stability broadband power calibrator, the instrument can output DC voltage, current and power, and has a dual voltage output mode, suitable for calibration power analyzer, AC / DC power meter, voltmeter, ammeter and other electrical measuring instruments.
 

2. Product features

  1. Standard Source of AC / DC Voltage: 0.3 V ~ 1020 V
  2. Standard Source of AC / DC Current: 0.5 mA ~ 20.5 A
  3. Standard Source of Auxiliary Voltage: 2.5 mV ~ 5.5 V
  4. Sine Wave Frequency: 5 Hz ~ 100 kHz
  5. Power Accuracy: Class 0.05
  6. Phase-Shifted Output
  7. Wideband Power Output
  8. A variety of quantity and value output mode
  9. Cross-guide amplifier (optional)

3. Main applications

☆ Calibration power analyzer (medium to low precision)
  1. The accuracy is Class 0.05, using the standard source method to calibrate the medium and low precision (Class 0.2 and below) power analyzer.
  2. The direct output of the maximum 20.5A, can output an current of 50A or 100A.
 
☆ Calibration power analyzer (medium to high precision)
  1. TA2100 matches TA3100 wide frequency power measurement standard with accuracy of 0.015%.
  2. Use the standard meter method to calibrate the power analyzer with medium and high precision (Class0.05 and below).

4. Main technical specifications

4.1 AC / DC Voltage Output

RangeResolving PowerVoltage Output RangeOutput Frequency RangeMaximum Load Current
3 V10 μV0.30000 V ~ 3.30000 VDC, 5 Hz ~ 100 kHz50 mA
10 V100 μV1.0000 V ~ 11.0000 VDC, 5 Hz ~ 100 kHz50 mA
30 V100 μV3.0000 V ~ 33.0000 VDC, 5 Hz ~ 100 kHz50 mA
100 V1 mV10.000 V ~ 110.000 VDC, 45 Hz ~ 100 kHz50 mA
300 V1 mV30.000 V ~ 330.000 VDC, 45 Hz ~ 100 kHz50 mA
600 V1 mV60.000 V ~ 660.000 VDC, 45 Hz ~ 100 kHz20 mA
1000 V10 mV100.00 V ~ 1020.00 VDC, 45 Hz ~ 10 kHz20 mA
 
RangeMeasurement Uncertainty at different frequencies (Hz) (k=2) @ (23°C ± 5°C), (ppm * reading + ppm * range)
DC5~4545~1k1k~10k10k~20k20k~50k50k~100k
3 V120+80300+200120+80300+200500+300600+4001000+500
10 V120+80300+200120+80300+200500+300600+4001000+500
30 V120+80300+200120+80300+200500+300600+4001000+500
100 V120+80-120+80300+200500+300600+4001000+500
300 V120+80-120+80300+200500+300600+4001000+500
600 V120+80-120+80300+200500+300600+4001000+500
1000 V120+80-120+80300+200---
  1. Range switching: manual or automatic shift
  2. Protection function: short-circuit protection, overload protection, overheat protection
 

4.2 AC / DC Current Output

RangeResolving PowerCurrent Output RangeOutput Frequency RangeMaximum Load Voltage
5 mA10 nA0.50000 mA ~ 5.50000 mADC, 5 Hz ~ 100 kHz7 V
10 mA100 nA1.0000 mA ~ 11.0000 mADC, 5 Hz ~ 100 kHz7 V
25 mA100 nA2.5000 mA ~ 27.5000 mADC, 5 Hz ~ 100 kHz7 V
50 mA100 nA5.0000 mA ~ 55.0000 mADC, 5 Hz ~ 100 kHz7 V
100 mA1 μA10.000 mA ~ 110.000 mADC, 5 Hz ~ 100 kHz7 V
200 mA1 μA20.000 mA ~ 220.000 mADC, 5 Hz ~ 100 kHz6 V
500 mA1 μA50.000 mA ~ 550.000 mADC, 5 Hz ~ 100 kHz6 V
1 A10 μA0.10000 A ~ 1.10000 ADC, 5 Hz ~ 100 kHz6 V
2.5 A10 μA0.25000 A ~ 2.75000 ADC, 5 Hz ~ 100 kHz6 V
5 A10 μA0.50000 A ~ 11.0000 ADC, 5 Hz ~ 100 kHz6 V
10 A100 μA1.0000 A ~ 11.0000 ADC, 5 Hz ~ 100 kHz6 V
20 A100 μA2.0000 A ~ 20.5000 ADC, 5 Hz ~ 100 kHz6 V
 
RangeMeasurement Uncertainty at different frequencies (Hz) (k=2) @ (23°C ± 5°C), (ppm * reading + ppm * range)
DC5~4545~1k1k~10k10k~20k20k~50k50k~100k
5 mA120+80300+200120+80300+200600+4001000+5002k + 1k
10 mA120+80300+200120+80300+200600+4001000+5002k + 1k
25 mA120+80300+200120+80300+200600+4001000+5002k + 1k
50 mA120+80300+200120+80300+200600+4001000+5002k + 1k
100 mA120+80300+200120+80300+200600+4001000+5002k + 1k
200 mA120+80300+200120+80300+200600+4001000+5002k + 1k
500 mA120+80300+200120+80300+200600+4001000+5002k + 1k
1 A120+80300+200120+80300+200600+4001000+5002k + 1k
2.5 A120+80300+200120+80300+200600+4001000+5002k + 1k
5 A120+80300+200120+80300+200600+4001000+5002k + 1k
10 A120+80300+200120+80300+200600+4001000+5002k + 1k
20 A120+80300+200120+80300+200600+4001000+5002k + 1k
  1. Range switch: manual or automatic shift; open circuit protection, overload protection, overheating protection

4.3 Auxiliary Voltage Output

RangeResolving PowerVoltage Output RangeOutput Frequency RangeMaximum Load Current
25 mV100 nV2.5000 mV ~ 27.5000 mVDC, 5 Hz ~ 100 kHz10 mA
50 mV100 nV5.0000 mV ~ 55.0000 mVDC, 5 Hz ~ 100 kHz10 mA
100 mV1 μV10.000 mV ~ 110.000 mVDC, 5 Hz ~ 100 kHz10 mA
250 mV1 μV25.000 mV ~ 275.000 mVDC, 5 Hz ~ 100 kHz10 mA
500 mV1 μV50.000 mV ~ 550.000 mVDC, 5 Hz ~ 100 kHz10 mA
1 V10 μV0.10000 V ~ 1.10000 VDC, 5 Hz ~ 100 kHz10 mA
2.5 V10 μV0.25000 V ~ 2.75000 VDC, 5 Hz ~ 100 kHz10 mA
5 V10 μV0.50000 V ~ 5.50000 VDC, 5 Hz ~ 100 kHz10 mA
 
RangeMeasurement Uncertainty at different frequencies (Hz) (k=2) @ (23°C ± 5°C), (ppm * reading + ppm * range)
DC5~4545~1k1k~10k10k~20k20k~50k50k~100k
25 mV150+200350+250150+200350+250600+4001000+5001200+800
50 mV150+200350+250150+200350+250600+4001000+5001200+800
100 mV150+200350+250150+200350+250600+4001000+5001200+800
250 mV150+200350+250150+200350+250600+4001000+5001200+800
500 mV120+80300+200120+80300+200500+300600+4001000+500
1 V120+80300+200120+80300+200500+300600+4001000+500
2.5 V120+80300+200120+80300+200500+300600+4001000+500
5 V120+80300+200120+80300+200500+300600+4001000+500
  1. Range switching: manual or automatic shift
  2. Wiring mode: BNC plug
  3. Protection function: short-circuit protection, overload protection, overheat protection
  4. Note: This output is especially suitable for the sensor small signal input of the power analyzer
 

4.4 Frequency

Adjustable rangeResolving PowerAnnual Measurement Uncertainty (k=2) @ (23°C ± 5°C)
5.000 0 Hz ≤ F ≤ 99.999 9 Hz0.000 1 Hz0.01%
100.000 Hz ≤ F ≤ 999.999 Hz0.001 Hz0.01%
1.000 00 kHz ≤ F ≤ 9.999 99 kHz0.01 Hz0.01%
10.000 0 kHz ≤ F ≤ 99.999 9 kHz0.1 Hz0.01%
100.000 kHz1 Hz0.01%
 

4.5 Phase

Phase Position[1][2]Annual Measurement Uncertainty (k=2) @ (23°C ± 5°C)
5~45Hz45~1kHz1k~10kHz10k~20kHz20k~50kHz50k~100kHz
φU -I0.05°0.02°0.05°0.1°0.2°0.5°
φU -AUX0.05°0.02°0.05°0.1°0.2°0.5°
Note [1]: Adjustment range: 0.000° ~ 359.999°
Note [2]: Resolution: 0.001°
Phase Position
(φ )
Power Factor
(λ )
The Active Power Uncertainty Component caused by the Phase Uncertainty[3]
5~45Hz45~1kHz1k~10kHz10k~20kHz20k~50kHz50k~100kHz
1.000000.000%0.000%0.000%0.000%0.001%0.004%
10°0.984810.015%0.006%0.015%0.031%0.062%0.158%
20°0.939690.032%0.013%0.032%0.064%0.128%0.321%
30°0.866030.050%0.020%0.050%0.101%0.202%0.508%
40°0.766040.073%0.029%0.073%0.147%0.294%0.736%
50°0.642790.104%0.042%0.104%0.208%0.417%1.044%
60°0.500000.151%0.060%0.151%0.302%0.605%1.515%
70°0.342020.240%0.096%0.240%0.480%0.960%2.401%
80°0.173650.495%0.198%0.495%0.990%1.980%4.953%
90°0.00000------
Note [3]: Calculation formula:
 

4.6  AC / DC Power Output

Quantity Range CombinationMeasurement Uncertainty at different frequencies (Hz) (k=2) @ (23°C ± 5°C), (% * power output)[1]
DC5~4545~1k1k~10k10k~20k20k~50k50k~100k
Voltage and Current Output (U & I) @ (λ =1)
(3 V … 30 V) &
(5 mA … 20 A)
0.030.070.030.070.130.180.34
(100 V … 600 V) &
(5 mA … 20 A)
0.03-0.030.070.130.180.34
1000 V &
(5 mA … 20 A)
0.03-0.030.07---
Dual Voltage Output (U & AUX) @ (λ =1)
(3 V … 30 V) &
(25 mV … 250 mV)
0.040.080.040.080.130.180.25
(3 V … 30 V) &
(500 mV … 5 V)
0.030.070.030.070.130.180.25
(100 V … 600 V) &
(25 mV … 250 mV)
0.04-0.040.080.130.180.25
(100 V … 600 V) &
(500 mV … 5 V)
0.03-0.030.070.130.180.25
1000 V &
(25 mV … 250 mV)
0.04-0.040.08---
1000 V &
(500 mV … 5 V)
0.03-0.030.07---
Note [1]: AC / DC power output range (virtual load): 0.15 mW ~ 20.91 kW
 Note [2]: Power factor adjustment range: -1 ~ 0 ~ + 1, resolution: 0.000 01
Note [3]: Power uncertainty calculation formula under other power factors: voltage and current output mode, dual voltage output mode, which: the voltage measurement uncertainty, the current measurement uncertainty, the auxiliary voltage measurement uncertainty, and the measurement uncertainty caused by the power factor.

 

5. General Technical Specifications

Power Supply VoltageAC (220±22) V ,(50±2) Hz
Preheating Time2 times from the last preheating time, up to 30 minutes.
Temperature PerformanceOperating temperature: 0°C~40°C
Calibration temperature: 18°C~28°C
Storage temperature: -20°C~70°C
Humidity PerformanceWorking Humidity: <80% @ 30°C, <70% @ 40°C
Storage humidity: (20%~80%) R · H, no condensation
Above Sea Class<3000 m
Communication InterfaceRS232×1,USB×1,LAN×1

Company Profile
 


Product Tag:
Related categories:
Scroll to Top