☆ Calibration for AC HV meter or Probe |
|
☆ Multiple Output Mode |
|
|
☆ Multiple Output Mode |
|
☆ Harmonics function | |
S/N | Function description |
1 | It can output 2nd~11th harmonic, and set harmonic amplitude. |
2 | Users can select the harmonics and set up harmonic parameter. |
3 | Oscilloscope function,display waveform of the output voltage. |
Range | Resolution | Stability (%/min) | Uncertainty (k=2) (ppm of reading + ppm of range) [1] | Distortion ( % ) | |
Class 0.05 | Class 0.02 | ||||
2 kV | 10 mV | 0.01 | 300 + 200 | 120 + 80 | ≤ 0.5 |
5 kV | 10 mV | 0.01 | 300 + 200 | 120 + 80 | ≤ 0.5 |
10 kV[2] | 100 mV | 0.01 | 300 + 200 | 120 + 80 | ≤ 0.5 |
15 kV[2] | 100 mV | 0.01 | 600 + 400 | 300 + 200 | ≤ 0.5 |
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). Note [2] : 10 kV,15 kV is optional. |
Power Supply | AC (220±22) V,(50 ± 2) Hz |
Temperature | Working temperature:0°C~45°C; Storage temperature:-20°C~70°C |
Humidity | Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity:(20%~80%) R·H,non-condensing |
Interface | RS232 interface |