PHILIPS MP70 Touch Screen ELO Part NO.D60370-000
PHILIPS MP70 Patient Monitor LCD Touch Screen ELO NO.D60370-000

Brief Description

PHILIPS MP70 touch screen ELO part NO.D60370-000 is the part on the front display screen of MP70 patient monitor, which is a glass touch screen of the display screen. Comes with a cable that connects to the main screen of the display, which can be used to connect the main screen of the MP70 patient monitor for normal use. This touch screen is a 15-inch color LCD, Up to eight waves can be shown on the screens, as well as the 12-Lead ECG Screen. It has a resolution of "4 hours @ 12 second resolution".

 

Detailed Photos
 
 
Function

PHILIPS MP70 touch screen is mainly used to display the data information on the MP70 patient monitor. Some main functions will be displayed on the main screen of the MP70 monitor first, and users can select the screen elements they need by touching or buttons. After selecting the function you need, the LCD screen will start to work. It can simultaneously display waveforms of eight vital signs, as well as use a 12-lead ECG. Very powerful.

 

More Deatils

Integrated Touch Display not functioning :

SymptomsPossible Causes of FailureFailure Isolation and Remedy
Touch Screen not
functioning
Flat panel adapter not connectedCheck connection from the flat panel adapter to the main board
Touch controller not connectedCheck connection from touch
controller to flat panel adapter
Touch panel cable not connectedCheck connection from touch
controller board to touch panel
Touchscreen functionality has been
permanently disabled
In service mode, select Main Setup -> User Interface and change the "Touch Enable" selection to "yes".
Touchscreen functionality has been
temporarily disabled
Check if touchscreen functioanlity has been temporarily disabled (padlock symbol on Main Screen key). If yes, press and hold the Main Screen key to re-enable touchscreen operation.
Main board defectiveReplace main board
Touch controller board or touch
sensor defective
Replace touch controller board and touch sensor
Note: Linearization data must be loaded and recalibrated after replacing the touch controller board
Touch Position invalidTouch not calibratedPerform touch calibration :
1. Enter the Main Setup Menu
2. Select Hardware
3. Select Touch Driver Settings
4. Select Calibrate

 