Model | NG-521 |
Glow wire | 4mm ± 0.04mm Ni/Cr(80/120) special standard ring, horizontally placed |
Thermocouple | 1.0mm K type, imported insulated high temperature armored thermocouple, armored sleeve heat resistant 1100 °C (better than standard 1050 °C) |
Heating temperature | 500 ºC ~ 960 ºC( could be preset ) ,temp fluctuation < 3 ºC |
Pressure | 0.95N±0.1N(can be set)(better than 1.0N±0.2N) |
Max depth | 7mm ± 0.5mm (adjustable) |
Sample moving speed | 10mm /s ~ 25mm /s |
Test time | 30s±0.1s (0.01s ~ 99.99s digital readout, preset ) |
Ignition plate thickness | 10mm thick white pine plate with size of 12g /m 2 ~ 30g /m 2 , standard silk paper, 200mm ± 5mm under the glow wire |
Test room | ≥ 0.5m³,black back, background illumination ≤ 20Lx |
Size | W 1150mm × D 600mm × H 1150mm, exhaust hole 100mm |
Power | 220V±10% 50HZ |