Biobase UV Vis Visible Double Beam Scanning Spectrophotometer

Min.Order: 1
Product origin: Jinan, Shandong, China
Infringement complaint: complaintComplaint
US$ 800 ~ 1300

Description

Product Description

Scanning UV/VIS Spectrophotometer BK-UV1800

Features:
1. Environmental lamp, preventing from Ozone inhalation.
2. Real-time monitoring the lifetime of lamp and Tungsten lamp.
3. Socket-Type lamp and Tungsten lamp.
4. GLP self-check function:check the wavelength accuracy and Photometry accuracy,provide test report and power-off protection.
5. Wavelength calibration, wavelength setting, change lamp source and dark current calibration automatically.
6. Si02 coating optical mirror, reduces the influence from outside.

 
 
 
 

Basic Functions:

1.Photometry

Test Abs., Transmittance and Energy by fixed wavelength

2. Quantitative

Linear fit and Linear fit through zero two modes

a. Coefficient, Standard Sample input and Standard Sample read three modes to establish standard curve 

b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings 

c. Can save 240 group curves, can test 240 data in each curve

d. Double wavelength, Triple wavelength test functions

3.Kinetics

Used for time course scanning or reaction rate calculations△A/t, can search all data.

4.Multi-Wavelength

Can test Transmittance and Abs. with 8 different wavelengths at most 

5.Scanning

User can set the scan range from 190nm to 1100nm to test the max. Abs. peak value, can do derivation, arithmetical operationsto the graph. 

6.Biology

6 methods: DNA/Protein, UV, Lowry, BCA,CBB and Biuret

Product Parameters

 

Model

BK-UV1800

BK-UV1800PC

BK-UV1600

BK-UV1600PC

BK-V1800

BK-V1800PC

BK-V1600

BK-V1600PC

Optical System

Single Beam, Grating 1200 lines/mm

Wavelength Range

190~1100nm

320-1100nm

Spectral Bandwidth

2nm

4nm

2nm

4nm

Wavelength Accuracy

± 0.5nm

Wavelength Repeatability

≤ 0.3nm

Photometric Accuracy

± 0.002A (0~0.5Abs), ± 0.004A (0.5~1.0Abs), ± 0.5% T (0~100% T)

Photometric Repeatability

0.001Abs (0~0.5Abs), 0.002Abs (0.5~1.0Abs), ≤ 0.2% T (0~100% T)

Stray Light

≤ 0.05% T @ 360nm; 220nm

≤ 0.05% T @ 360nm;220nm

Stability

± 0.001A / h @ 500nm

Baseline Flatness

± 0.002A

Noise

± 0.0005A

Display

480 * 272 65 thousand true color TFT LCD

Photometric Mode

T, A, C, E

Photometric Range

0~200% T, -0.301~3.0A

Detector

Silicon Photodiode

Light Source

Lamp, Tungsten Lamp

Input

Membrane Keypad

Output

USB  Print and data output, Connect PC

Standard Accessories

10mm glass cuvette *4 units      10mm quartz cuvette *2 units

10mm glass uvette *4 units

Optional Accessories

Single cell holder,General cell holder,Test tube holder,Solid sample holder, lamp,Halogen lamp,4-place Cell holder,Printer,Water-jacketed cell holder,Reflection accessory 5°,Auto 8-cell changer,Micro cell holder,Water-jacketed cell holder,Constant-temperature system

 

4-place Cell holder,Printer, Test tube holder, lamp,Halogen lamp

 

Power Supply

AC110V/220±10% ,60/50Hz

Packing Size(mm)

620*500*370

620*500*370

620*500*370

620*500*370

Gross Weight (kg)

18

18

16.5

16.5

 
 

 