Hospital ICU Machine Portable Monitor Medical Patient Monitoring System E12

Min.Order: 5
Product origin: Shanghai, China
Infringement complaint: complaintComplaint
US$ 500 ~ 600

Description
                   Medical Products Emergency Room Professional Portable Patient Monitor E12

For all E series modular monitor:
Standard Config: 3/5 lead ECG.RESP.SpO2,PR,NIBP,2-Temp, Lithium Battery, Touch Screen
Option Module:2-IBP,Maslimo Spo2,Sidestream/Microflow/Mainstream EtCo2,Mainstream /Sidestream C.O.
Other option:9 lead ECG, Printer, Rolling stand, Wall mount, VGA, WIFI,SD Memory card,Touch Screen

WIFI with smart IT solutions
1.Wireless integration with Central Monitoring Station
2.Dynamic trends provide up to 240 hours of usefulinformation for review
3.8 traces per monitor and 16 monitors on one screen
4.View up to 32 maximum bed on one platform in real-time
5.Review and manage patient data anytime and anywhere in and pre-hospita


Specification
ModelMedical ICU Monitor E12
Display 12.1 real color TFT optional touch screen
Operating environment: Temperature0~40ºC
Humidity-≤85%
Altitude:-500 ~ 4600m
Transport and Storage environmentTemperature:-20~6OºC
Humidity≤93%
Altitude:-500~13100m
Power Requirements AC/DC100~-240V,50Hz/60Hz / Built-in rechargeable battery
Battery4000mA 11.1V lithium battery
2h operating after full charge(one piece)
5min operating after low battery alarm
ST Segment Range-2.0mV ~ +2.0mV
ECG 5 LeadsRA,LA,LL,RL,V
Packing380*350*300mm
Weight6.1KG

Date Storage
Trend diagram/table:7*24h
NIBP review:400 events
Wave review:100min
Alarm review:100 alarm events
Support drug concentration titration analysis

ST Segment
ST Segment Range:-2.0mV ~ +2.0mV
Accuracy:0.02mv

ECG
5 Leads: RA,LA,LL,RL,V
Lead mode:1.li.II, aVR,aVL.aVF.V
Increase: *250,*500,*1000,*2000
Sensitivity: > 200 uV(Peak-to-pealk value)
Input impedance: > 5( megohm )
Bandwidth: Surgery 1~ 20Hz
Monitor 0.5~40Hz
Diagnostic 0.05~130 Hz
CMRR:>=100dB
Polarization voltage:+- 300mv
Baseline Recowering Time: After defibrillation< 3 seconds
Signal Range: 8 mV CPeak-to-peak value)
Calibrating Signal:1mV , precision +-5%

RESP
Method: RA-LL impedance
Resp lmpedance range:0.3~3 0
Base lmpedaince range:200Ω~4000Ω 
Bandwidth:0.1~2.5Hz
Resp rate: Adult                0~120BrPM_
         NeonatallPediatric   0~150BrPM
         Resolution:                            1BrPM
         Precision:                               +-2 BrPM
Asphyxia Alarm: 10~40 seconds

NIBP
Method;Pullse wave oscillometry
Work mode:ManuallAuto/ STAT
Measure interval of auto mode:
1.2,3.4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode: 5 minutes
PR range.40~240bpm
Measure & alarm range:      
 Adult
SYS40~ 270mmHg DIA10~ 215mmHg MEAN20~ 235mmHg
Pediatric
SYS40~200mmHg DIA10~150mmHg MEAN20~165mmlHg
Neonatal
SYS40~135mmHg DIA10~ 100mmHg MEAN20~110mmHg
Static pressure range:  0~ 300mmHg
                     Precision: +- 3mmHg
Pressure precision:   Max. awverage error: +-5mmHg
                     Ma.standard deviation: +-8mmHlg
Overvoltage protection:   Adult 300mmHg
                         Pediatric 240mmHg
                         Neonatal 150mmHg






 

FAQ :

1.If your company support OEM ?
Yes, we support OEM, You just sent me our your logo and requirement, we will print the logo in the products and nameplate.

2.What's our payment ways and methods?
Our payment ways is T/T: 30% before producing, and 70% before shipping the products.
L/C at sight and west union. 

3.What is terms of the transaction ?
Our general quotation is EXW, and we will quote CFR after confirming detailed quantity with customer. We also support DDP.

4.What is your delivery time?
For small pcs such as 1 pcs , the exact delivery time is about 5 days after receiving the payment.
For large pcs,such as 100 pcs, about one month after receiving the advance payment.

               








 
Product Tag:
Related categories:
Scroll to Top