Features:
1.Large LCD screen (128*64 Dots)with saving standard curvr function
2.Imported&Environmental Deuteriu lamp,preventing from Ozone inhalation
3.Adopted high-class grating with wholly heremetic light path design,to ensure the instrument has the super low stray light.
4.Real-time monitoring the lifetime of Deuteriu lamp and Tungsten lamp with advanced system
5.Pre-aligned design ensures the user can change lamps conveniently.
6.Wavelength calibration,wavelength setting,change lamp conveniently.
7.With SiO2 coating optical mirror,reducing the pollution from outside fully
8.Optional PC Software to expand the applications to Quantitative,Multi-Wavelength and Kinetics.
9.Widely used for Organic chemistry,Inorganic chemistry,Life sciences,foods,Medicine and health ,Agriculture,Geology,Metallurgy and Environment etc. fields.
Main Technical Parameters
Model | FUV-1000 | FV-1000 | FUV-1200 | FV-1200 |
Optical System | Single Beam,Grating 1200 lines/mm | |||
Wavelength Range | 200- 1020nm | 325- 1000nm | 200- 1020nm | 320- 1020nm |
Spectral Bandwidth | 4nm | |||
Wavelength Accuracy | ±1.8nm | ±1.0nm | ||
Wavelength Repeatability | ≤0.5nm | |||
Photometric Accuracy | ±0.002A(0-0.5Abs), ±0.004A(0.5-1.0Abs), ±0.5%T(0-100% T) | |||
Photometric Repeatability | 0.001Abs(0-0.5Abs), 0.002Abs(0.5-1.0Abs), ≤0.2%T(0-100% T) | |||
Stray Light | ≤0.002%T@360nm; 220nm (220nm for UV only) | |||
Stability | ±0.002A/h@500nm | |||
Noise | 0.001A | |||
Display | 128*64 Dots LCD | |||
Photometric Mode | T,A,C,E | |||
Photometric Range | 0-200% T,-0.301-3.0A | |||
Detector | Silicon Photodiode | |||
Light Source | Deuteriu Lamp,Tungsten Lamp | |||
Input | Membrane Keypad | |||
Output | USB-A,USB-B | |||
Compartment | Optional Solid Sample Holder, Micro Cell Holder, 100-100mm Cell Holder |