Portable aerosol generator according to the principle of occurrence can be divided into condensation and the occurrence of two types of heat generation. Among them, the condensation of the aerosol generation of AG-220 (with pump), AG-230 (compressed air type), heating aerosol generation is AG-1800 type.
Aerosol generator can be polydisperse particles, in the pressure of 20pisg and flow rate of 50 ~ 5400cfm range, can occur at a concentration of 10 ~ 100ug / L aerosol particles (AG-230).
Aerosol generator that produces smoke equipment, DOP solvent (or PAO solvent) into the generator container, under a certain pressure or heating conditions can produce aerosol smoke, the aerosol smoke into the high efficiency filter on the wind side (DOP liquid heating, the formation of DOP steam, steam under certain conditions, condensed into tiny droplets, remove too large and too small droplets, leaving only about 0.3μm particles, mist DOP into the duct).
High efficiency filter and its installation If there are defects, such as the filter itself has a small hole or the installation does not form a tight crack, will lead to less than the intended purification effect. Therefore, after the installation or replacement of the high-efficiency filter, the filter and the installation connection must be leaked. By testing the leaks of the high efficiency filter, it is found that the high efficiency filter and its installation are flawed in order to take remedial action.
Aerosol generator and aerosol photometer is used to detect high efficiency filter leakage efficiency. The aerosol generator is used in combination with an aerosol photometer, and defects such as pinholes, leaks, gasket leaks, leaks, or frame leaks can be quickly detected and corrected in a timely manner , Thus ensuring efficient filter efficiency and cleanliness between the cleanliness level.
AG-220, AG-230 advantages In addition to stainless steel, a larger filling port, independent Laskin Nozzles control, and the occurrence of each generator concentration and applicable flow range is not the same, the user according to different needs Arbitrary choice.
Model / Parameter | AG 230 (compressed air type) | AG 220 (virith pump type) | AG-1800 (heat generation type) |
Available flow rates range | 50-5400cfm(1.4-150m³/min) | 50-2000)cfm(1.4-55m³/min) | 1500-65000cfm(42-1800m³/min) |
Suspended particle concentration | 10ug/L: airflow5400rfrn (about 150m³/min) | 10ug/L: a ir flow 2000cfm(about 55m³/min) | 10ug/L airflow 65000cfm (about 1800m³/ min) |
Suspended particle concentration | 100ug/L: airflow 540cfm (about 15m³/min) | 100ug/L: airflow 200cfm (about 5.5m³/min) | 10Oug/L: airflow 6500cfm (about180m³/min) |
The occurrence of particles | DOP PAO , Polydisperse | ||
Method of occurrence | 2-4 Laskin Nozzles | 2-6 Laskin nozzles | |
Compressed gas | 3-12cfm(85-340Umin)20pis(0.14Mpa) | Built-in compressed air machine | Nitrogerv Argorv Carborv Dioxide or Helium (50psi) |
Power supply | No need | 220VAC. 50Hz | |
Outer dimension | 220 Wx 280 D x300 H(mm) | 200 Wx 270 D x550 H(mm) | 350 Wx 250D x250 H(mm) |
Weight | 6.5kg | 25kg | 9.1kg |